Aflatoxins B1 Rapid Test Strips

【Instruction】                                                       

     The product is used for the rapid detection residue of aflatoxin B1             

 (AFLB1) in Corn,wheat etc. Grain and feed sample.The total detection

 time is 5 minutes and the detection limit is 20ppb.Applicable to all types 

of enterprises and institutions.

 

【Components】

1.Aflatoxin B1 Rapid Test Strips( 40pcs/box)

2.Dropper(40pcs/box)

3.Desiccant

4Instruction(1pc/box)

 

【Notes】

1.Wear gloves when test

2.Every test cassette and dropper are single use only. Please do not reuse.

3.Storage: store at 4-30°C & dry place keep from light ,No frozen.

 

【Expiry date】

12 months; date of production is on box.

 

【Sample Preparation】

1.Taking 2.0g Grinded corn samples in 5ml centrifuge tube(The corn flour  

  is not suitable for the detection).

2Adding 3ml 20% methanol water(methanol/water=1:4),Severe reverse

  shake for 10 seconds. Then be static for 10 seconds. Waiting for test.

【Operation Procedure】

Read the instruction carefully before use. Return the test strip and inspected sample solution to room temperature (20 ℃ ~ 30 ℃) before test.

 

1. Taking out the test strip from the original package (immediately using after opening).Put the  

   test strip flatly , show on the right side of the observer, as shown below:

                      

2. Using Pipette to absorb 150 microliters of sample solution ( or Draw 5 drop dropper) .Adding

   it to sample collect region.

3. Reading the result in 3-5 minutes. Other time the result is invalid.

 

Negative(-):"T' line is darker than" C" line or same thick color.Said  aflatoxin B1 concentrations of the sample below the detection limit or no aflatoxin B1 residues.
Positive(+):"T' line is lighter than" C" line or "T" line is without color.Said in a sample of aflatoxin B1 

concentrations more than the detection limit. If "T" line is much lighter than "C" line or without color .

Said that the concentrations of aflatoxin B1 from sample is much high.
Invalid: No band appears in C zone.

 